D8 Venture SC-XRD

D8 Venture SC-XRD

Instrument specification

Bruker Single Crystal X-ray Diffractometer (SCXRD), D8 Venture with Photon III C14 detector, is used to determine three dimensional crystal structures. It is equipped with two micro-focus X-ray sources (Cu and Mo radiation) providing a choice of two wavelengths, for measuring patterns in very small crystals. Powder diffraction of capillary samples can also be obtained. Grazing incidence measurements can also be made. Different cooling systems are available, enabling low temperatures such as 100 K.

D8 Venture Single Crystal X-Ray Diffractometer

Technicalities

Bruker SCXRD D8 Venture provides a high intense strongest X-ray source, with highly reliable largest X-ray detectors. With more room for rotating anode, liquid metal jet, and dual wavelength solutions, this instrument provides high versatility and flexibility to the user. The IμS Microfocus Sources yield Mo 0.71 Å and Cu 1.54 Å. The Mo radiation helps in high throughput and electron density measurements, and the Cu radiation provides absolute configuration determination of lightweight molecules.

Key features of D8 Venture SC-XRD

Theory of operation

A crystal has a pattern of unit cells repeated in a three-dimensional lattice. Upon X-rays incident on the crystal lattice, it behaves like a diffraction grating, producing a diffraction pattern. A beam of parallel monochromatic X-rays of a specific wavelength is used to obtain this diffraction. The analysis of this diffraction pattern leads to the structure and shape of the lattice units. With the phase angles and a good fit between experimental and calculated diffraction patterns.